The future of motion artefacts correction in MRI

Authors

  • Dubravko Bobinec Children’s Hospital Zagreb, The Department of Pediatric Radiology Author
  • Krešimir Dolic University Hospital of Split , University Department of Health Studies, Split, Croatia Author

DOI:

https://doi.org/10.55378/rv.47.2.2

Keywords:

MRI, motion artefact, methods, future, sequences, correction

Abstract

Motion during the MR examination affects the degradation of the image in the form of „ghosting“ or blurring and is one of the last major unresolved issues. The patient’s movements can be of different scope, directions and orientation, making it difficult to find a solution from the early days of MR (Magnetic Resonance) development.

Today, there are three main strategies for combating motion artefacts: motion prevention, motion reduction and motion correction. Due to the lack of possibilities and exhausted ideas, most work is being done on the development of motion correction using MR navigators or prospective and retrospective methods.

Of the future techniques, the most promising are SMS (Simultaneous Multi-Slice) sequences and CS (Compres­sed Sensing) due to the resistance to movement and shorter recording time compared to conventional sequences.

There is currently no general solution for the artefact of motion, but the problem is approached according to a specific case by choosing the method with the least shortcomings and taking into account the type of movement and the sequence of selection. More research is needed in clinical settings given the complexity and diversity of the problem with the involvement of a multidisciplinary team.

References

Sadigh G, Applegate KE, Saindane AM. Prevalence of unanticipated events associated with MRI examinations: a benchmark for MRI quality, safety, and patientexperience. J Am Coll Radiol. 2017;14:765–772.

Gumus K, Keating B, White N, et al. Comparison of optical and MR-based tracking. Magn Reson Med. 2015;74:894–902.

Zaitsev M, Maclaren J, Herbst M. Motion artifacts in MRI: A complex problem with many partial solutions. J Magn Reson Imaging. 2015 Oct;42(4):887-901.

Brown TT, Kuperman JM, Erhart M, et al. Prospective motion correction of high-resolution magnetic resonance imaging data in children. Neuroimage 2010;53:139–145.

Thesen S, Heid O, Mueller E, Schad LR. Prospective acquisition correction for head motion with image-based tracking for real-time fMRI. Magn Reson Med 2000;44:457–463.

Zaitsev M, Dold C, Sakas G, Hennig J, Speck O. Magnetic resonance imaging of freely moving objects: prospective real-time motion correction using an external optical motion tracking system. Neuroimage 2006;31:1038–1050.

Ooi MB, Krueger S, Thomas WJ, Swaminathan SV, Brown TR. Prospective real-time correction for arbitrary head motion using active markers. Magn Reson Med 2009;62:943–954.

Fu ZW, Wang Y, Grimm RC, et al. Orbital navigator echoes for motion measurements in magnetic resonance imaging. Magn Reson Med 1995;34:746–753.

Welch EB, Manduca A, Grimm RC, Ward HA, Jack CR JR. Spherical navigator echoes for full 3D rigid body motion measurement in MRI. Magn Reson Med 2002;47:32–41.

Costa AF, Petrie DW, Yen YF, Drangova M. Using the axis of rotation of polar navigator echoes to rapidly measure 3D rigid-body motion. Magn Reson Med 2005;53:150–158.

Qin L, Gelderen P, Zwart J, Jin F, Tao Y, Duyn J. Head movement correction for MRI with a single camera. In: Proceedings of the 16th Annual Scientific Meeting of ISMRM, Toronto, Ontario, 2008. (abstract 1467).

Korin HW, Felmlee JP, Riederer SJ, Ehman RL. Spatial-frequencytuned markers and adaptive correction for rotational motion. Magn Reson Med 1995;33:663–669.

Sengupta S, Tadanki S, Gore JC, Welch EB. Prospective real-time head motion correction using inductively coupled wireless NMR probes. Magn Reson Med 2014;72:971–985.

Maclaren J, Armstrong BS, Barrows RT, et al. Measurement and correction of microscopic head motion during magnetic resonance imaging of the brain. PLoS One 2012;7:e48088.

Schulz J, Siegert T, Reimer E, et al. An embedded optical tracking system for motion-corrected magnetic resonance imaging at 7T. Magn Reson Mater Phys 2012;25:443–453.

Feinberg DA, Giese D, Bongers DA, et al. Hybrid ultrasound MRI for improved cardiac imaging and real time respiration control. Magn Reson Med 2010;63:290–296.

Speck O, Hennig J, Zaitsev M. Prospective real-time slice-by-slice motion correction for fMRI in freely moving subjects. Magn Reson Mater Phys 2006;19:55–61.

Schulz J, Siegert T, Bazin PL, et al. Prospective slice-by-slice motion correction reduces false positive activations in fMRI with taskcorrelated motion. Neuroimage 2014;84:124–132.

Gumus K, Keating B, Poser BA, et al. Prevention of motion-induced signal loss in diffusion-weighted echo-planar imaging by dynamic restoration of gradient moments. Magn Reson Med 2014;71:2006–2013.

Herbst M, Maclaren J, Weigel M, Korvink J, Hennig J, Zaitsev M. Prospective motion correction with continuous gradient updates in diffusion weighted imaging. Magn Reson Med 2012;67:326–338.

Keating B, Deng W, Roddey JC, et al. Prospective motion correction for single-voxel 1H MR spectroscopy. Magn Reson Med 2010;64:672–679.

Zaitsev M, Speck O, Hennig J, Buchert M. Single-voxel MRS with prospective motion correction and retrospective frequency correction. NMR Biomed 2010;23:325–332.

Lange T, Maclaren J, Buechert M, Zaitsev M. Spectroscopic imaging with prospective motion correction and retrospective phase correction. Magn Reson Med 2012;67:1506–1514.

Boegle R, Maclaren J, Zaitsev M. Combining prospective motion correction and distortion correction for EPI: towards a comprehensive correction of motion and susceptibility-induced artifacts. Magn Reson Mater Phys 2010;23:263–273.

Atkinson D, Hill DLG, Stoyle PNR, et al. Automatic compensation of motion artifacts in MRI. Magn Reson Med 1999;41:163–170.

McGee KP, Felmlee JP, Jack CR Jr, Manduca A, Riederer SJ, Ehman RL. Autocorrection of three-dimensional time-of-flight MR angiography of the Circle of Willis. AJR Am J Roentgenol 2001;176:513–518.

Loktyushin A, Nickisch H, Pohmann R, Scholkopf B. Blind retrospective motion correction of MR images. Magn Reson Med 2013;70:1608–1618.

Lauzon ML, Rutt BK. Generalized K-space analysis and correction of motion effects in MR imaging. Magn Reson Med 1993;30:438–446.

Beatty PJ, Nishimura DG, Pauly JM. Rapid gridding reconstruction with a minimal oversampling ratio. IEEE Trans Med Imaging 2005;24:799–808.

Pruessmann KP, Weiger M, Bornert P, Boesiger P. Advances in sensitivity encoding with arbitrary k-space trajectories. Magn Reson Med 2001;46:638–651.

Atkinson D, Hill DL. Reconstruction after rotational motion. Magn Reson Med 2003;49:183–187.

Marxen M, Marmurek J, Baker N, Graham SJ. Correcting magnetic resonance k-space data for in-plane motion using an optical position tracking system. Med Phys 2009;36:5580–5585.

Leung G, Plewes DB. Retrospective motion compensation using variable-density spiral trajectories. J Magn Reson Imaging 2005;22:373–380.

Aksoy M, Forman C, Straka M, Cukur T, Hornegger J, Bammer R. Hybrid prospective and retrospective head motion correction to mitigate cross-calibration errors. Magn Reson Med 2012;67:1237–1251.

Schmidt JF, Wissmann L, Manka R, Kozerke S. Iterative k-t principal component analysis with nonrigid motion correction for dynamic three-dimensional cardiac perfusion imaging. Magn Reson Med 2014;72:68–79.

Schmidt JF, Buehrer M, Boesiger P, Kozerke S. Nonrigid retrospective respiratory motion correction in whole-heart coronary MRA. Magn Reson Med 2011;66:1541–1549.

Bydder M, Larkman DJ, Hajnal JV. Detection and elimination of motion artifacts by regeneration of k-space. Magn Reson Med 2002;47:677–686.

Wintersperger BJ, Runge VM, Biswas J, et al. Brain magnetic resonance imaging at 3 Tesla using BLADE compared with standard rectilinear data sampling. Invest Radiol. 2006;41:586–592.

Shin HJ, Kim MJ, Lee MJ, et al. Comparison of image quality between conventional VIBE and radial VIBE in free-breathing paediatric abdominal MRI. Clin Radiol. 2016;71:1044–1049.

Runge VM, Richter JK, Heverhagen JT. Motion in Magnetic Resonance: New Paradigms for Improved Clinical Diagnosis. Invest Radiol. 2019 Jul;54(7):383-395.

Runge VM, Clanton JA, Partain CL, et al. Respiratory gating in magnetic resonance imaging at 0.5 Tesla. Radiology. 1984;151:521–523.

Liu YL, Riederer SJ, Rossman PJ, et al. A monitoring, feedback, and triggering system for reproducible breath-hold MR imaging. Magn Reson Med. 1993;30:507–511.

Feng L, Axel L, Chandarana H, et al. XD-GRASP: golden-angle radial MRI with reconstruction of extra motion-state dimensions using compressed sensing. Magn Reson Med. 2016;75:775–788.

Taron J, Weiss J, Notohamiprodjo M, et al. Acceleration of magnetic resonance cholangiopancreatography using compressed sensing at 1.5 and 3 T: a clinical feasibility study. Invest Radiol. 2018;53:681–688.

Yoon JH, Lee SM, Kang HJ, et al. Clinical feasibility of 3-dimensional magnetic resonance cholangiopancreatography using compressed sensing: comparison of image quality and diagnostic performance. Invest Radiol. 2017;52:612–619.

Zhu L, Wu X, Sun Z, et al. Compressed-sensing accelerated 3-dimensional magnetic resonance cholangiopancreatography: application in suspected pancreatic diseases. Invest Radiol. 2018;53:150–157.

Wetzl J, Schmidt M, Pontana F, et al. Single-breath-hold 3-D CINE imaging of the left ventricle using Cartesian sampling. MAGMA. 2018;31:19–31.

Naresh NK, Haji-Valizadeh H, Aouad PJ, et al. Accelerated, first-pass cardiac perfusion pulse sequence with radial k-space sampling, compressed sensing, and k-space weighted image contrast reconstruction tailored for visual analysis and quantification of myocardial blood flow. Magn Reson Med. 2019;81:2632–2643.

Bustin A, Ginami G, Cruz G, et al. Five-minute whole-heart coronary MRA with sub-millimeter isotropic resolution, 100% respiratory scan efficiency, and 3DPROST reconstruction. Magn Reson Med. 2019;81:102–115.

Bacher M, Speier P, Bollenbeck J, et al. Pilot tone navigation enables contactless prospective cardiac triggering: initial volunteer results for prospective cine. Proceedings of the 26th Annual Meeting of the ISMRM, #2960. 2018. [citirano 15.3.2022.]. Dostupno na: http://archive.ismrm.org/2018/2960.html.

Setsompop K, Gagoski BA, Polimeni JR, et al. Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g-factor penalty. Magn Reson Med. 2012;67:1210–1224.

Barth M, Breuer F, Koopmans PJ, et al. Simultaneous multislice (SMS) imaging techniques. Magn Reson Med. 2016;75:63–81.

Runge VM, Richter JK, Heverhagen JT. Speed in clinical magnetic resonance. Invest Radiol. 2017;52:1–17.

Setsompop K, Cohen-Adad J, Gagoski BA, et al. Improving diffusion MRI using simultaneous multi-slice echo planar imaging. Neuroimage. 2012;63:569–580.

Cauley SF, Polimeni JR, Bhat H, et al. Interslice leakage artifact reduction technique for simultaneous multislice acquisitions. Magn Reson Med. 2014;72:93–102.

COVER The future of motion artefacts correction in MRI

Downloads

Published

2023-05-16

How to Cite

Bobinec, Dubravko, and Krešimir Dolic. 2023. “The Future of Motion Artefacts Correction in MRI”. Radiological Journal 47 (2): 10-17. https://doi.org/10.55378/rv.47.2.2.

Similar Articles

1-10 of 51

You may also start an advanced similarity search for this article.

Most read articles by the same author(s)