Influence of vertical position of the table from the isocenter on radiation dose in CT imaging

Authors

DOI:

https://doi.org/10.55378/rv.50.1.4

Keywords:

CT, isocenter, CTDIvol, automatic exposure control (AEC), dose optimization

Abstract

Precise positioning of the patient at the CT scanner isocenter is essential for achieving optimal image quality and minimizing radiation dose. Vertical deviations from the isocenter can alter projection geometry on the scout image, affecting the performance of the automatic exposure control (AEC) system and leading to unintended dose variations. Despite its importance, incorrect patient centering remains common in routine clinical practice.

The aim of this study was to quantify the effect of vertical table height deviation relative to the isocenter on CTDIvol and DLP during CT imaging of thoracic and pelvic phantoms. Measurements were performed on two CT scanners, GE Revolution Maxima 128 and GE Revolution ES 256. The table height was adjusted from –10 cm to +10 cm in 2 cm increments relative to the isocenter. Imaging was conducted using a thoracic phantom and an anthropomorphic pelvic phantom with standardized and optimized protocols specific to each scanner. Two scans were performed at each table position to ensure measurement reliability, and dose metrics were extracted from the dose reports.

Results showed that raising the table above the isocenter caused a substantial increase in radiation dose on both scanners. On the GE Revolution Maxima, a +10 cm deviation led to dose increases of 71% for thoracic imaging and 106% for pelvic imaging, whereas lowering the table reduced the dose by up to 30%. On the GE Revolution ES, the maximum dose increase was 53%, with dose reductions of up to 20% when the table was lowered. Dose variations were more pronounced in pelvic protocols, likely due to greater phantom thickness and increased AEC sensitivity.

In conclusion, vertical miscentering has a significant and clinically relevant impact on CT radiation dose. Accurate patient positioning at the isocenter is a simple yet critical dose optimization measure in CT practice.

 

References

International Commission on Radiological Protection. Radiological protection in medicine. ICRP Publication 105. Ann ICRP. 2007;37(6):1-63.

Kalra MK, Maher MM, Toth TL, Hamberg LM, Blake MA, Shepard JA, et al. Strategies for CT radiation dose optimization. Radiology. 2004;230(3):619-628.

American Association of Physicists in Medicine (AAPM). Size-Specific Dose Estimates (SSDE) in Pediatric and Adult Body CT Examinations. AAPM Report No. 204.

Li J, Udayasankar UK, Toth TL, Seamans J, Small WC, Kalra MK. Automatic patient centering for MDCT: effect on radiation dose. AJR Am J Roentgenol. 2007;188(2):547-552.

Kaasalainen T, Palmu K, Reijonen V, Kortesniemi M. Effect of patient centering on patient dose and image noise in chest CT. AJR Am J Roentgenol. 2014;203(1):123-130.

Kim MS, Singh S, Kalra MK, Halpern EF, Saini S. Relationship between patient centering, mean CT numbers and noise in abdominal CT: influence of anthropomorphic parameters. World J Radiol. 2012;4(3):102-108.

Toth T, Ge Z, Daly MP. The influence of patient centering on CT dose and image noise. Med Phys. 2007;34(7):3093-3101.

Habibzadeh MA, Ay MR, Kamali Asl AR, Ghadiri H, Zaidi H. Impact of miscentering on patient dose and image noise in x-ray CT imaging: phantom and clinical studies. Phys Med. 2012;28(3):191-199.

Barreto I, Lamoureux R, Olguin C, Quails N, Correa N, Rill L, et al. Impact of patient centering in CT on organ dose and the effect of using a positioning compensation system. J Appl Clin Med Phys. 2019;20(6):141-151.

Huda W, Mettler FA. Volume CT dose index and dose-length product displayed during CT: what good are they? Radiology. 2011;258(1):236-242.

Naslovnica Utjecaj vertikalnog odstupanja od izocentra na dozu zračenja u CT dijagnostici

Downloads

Published

2026-05-05

Issue

Section

Original scientific papers

How to Cite

Krpan, Marko, Bruno Rajic, Marin Brajkovic, Vedran Markotic, Ivan Lasic, and Ante Bosnjak. 2026. “Influence of Vertical Position of the Table from the Isocenter on Radiation Dose in CT Imaging”. Radiological Journal 50 (1): 17-23. https://doi.org/10.55378/rv.50.1.4.

Similar Articles

1-10 of 38

You may also start an advanced similarity search for this article.