Doze zračenja na različitim digitalnim sustavima u radiologiji

Autor(i)

  • Goran Bajs Dom zdravlja Zagreb Istok , University of Split, Faculty of Health Sciences, Split, Croatia Autor
  • Dario Posedel University of Applied Health Sciences Zagreb, Zagreb, Croatia
  • Frane Mihanović University of Split, Faculty of Health Sciences, Split, Croatia

DOI:

https://doi.org/10.55378/rv.50.2.3

Ključne riječi:

računalna radiografija, direktna digitalna radiografija, doza zračenja, ulazna površinska doza zračenja (ESD), radiološka zaštita

Sažetak

Uvod: Ovaj rad analizira doze zračenja na digitalnim radiografskim sustavima računalne radiografije (CR) i direktne digitalne radiografije (DR) pri snimanju pluća, ramena i zdjelice s kukovima.

Cilj: Cilj istraživanja bio je usporediti ulazne površinske doze zračenja (ESD) između CR i DR sustava te procijeniti mogućnosti DR tehnologije u optimizaciji doze zračenja.

Metode: Istraživanje je provedeno na 180 ispitanika različitog spola i životne dobi. ESD je izračunat na temelju podataka dobivenih godišnjim ispitivanjem rendgenskih uređaja iz 2025. godine. Za usporedbu CR i DR sustava primijenjen je t-test za nezavisne uzorke, za analizu razlika između muškaraca i žena Mann–Whitney U test, a linearna regresijska analiza korištena je za procjenu povezanosti vrijednosti dobivenih na CR i DR sustavima. Statistička značajnost postavljena je na p < 0,05.

Rezultati: DR sustav ostvario je niže srednje vrijednosti ESD-a u svim analiziranim anatomskim regijama. Statistički značajne razlike utvrđene su kod snimanja pluća i ramena, dok kod zdjelice razlika nije bila statistički značajna. Regresijska analiza nije pokazala značajnu linearnu povezanost između vrijednosti ESD-a dobivenih na CR i DR sustavima.

Zaključak: U uvjetima ovog istraživanja DR sustav pokazao je niže vrijednosti ESD-a u odnosu na CR sustav, osobito pri snimanju pluća i ramena. Dobiveni rezultati ističu važnost pravilne primjene AEC sustava i optimizacije ekspozicijskih parametara u cilju smanjenja izloženosti pacijenata ionizirajućem zračenju.

Reference

Bushberg JT, Seibert JA, Leidholdt EM Jr, Boone JM. The Essential Physics of Medical Imaging. 4th ed. Philadelphia: Wolters Kluwer; 2020.

International Atomic Energy Agency. Dosimetry in Diagnostic Radiology: An International Code of Practice. Technical Reports Series No. 457. Vienna: IAEA; 2007.

Carlton RR, Adler AM, Balac V. Principles of Radiographic Imaging: An Art and A Science. 6th ed. Boston: Cengage Learning; 2020.

Jones AK, Heintz P, Geiser W, Goldman L, Jerjian K, Martin M, et al. Ongoing quality control in digital radiography: Report of AAPM Imaging Physics Committee Task Group 151. Med Phys. 2015;42(11):6658–6670. doi:10.1118/1.4932623.

Sonoda M, Takano M, Miyahara J, Kato H. Computed radiography utilizing scanning laser stimulated luminescence. Radiology. 1983;148(3):833-838.

Seeram E. Digital Radiography: Physical Principles and Quality Control. 2nd ed. Cham: Springer; 2019.

Seibert JA. Flat-panel detectors: how much better are they? Pediatr Radiol. 2006;36(Suppl 2):173-181.

International Electrotechnical Commission. IEC 62220-1-1: Medical electrical equipment – Characteristics of digital X-ray imaging devices – Determination of detective quantum efficiency. Geneva: IEC; 2015.

International Commission on Radiological Protection. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP. 2007;37(2-4):1-332.

Lin P-P, Schueler BA, Balter S, Strauss KJ, Wunderle KA, LaFrance MT, et al. Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: a report of the AAPM Imaging Physics Committee Task Group 190. Med Phys. 2015;42(12):6815–6829. doi:10.1118/1.4934831.

International Atomic Energy Agency. Diagnostic Radiology Physics: A Handbook for Teachers and Students. Vienna: International Atomic Energy Agency; 2014.

United Nations Scientific Committee on the Effects of Atomic Radiation. Sources, Effects and Risks of Ionizing Radiation: UNSCEAR 2020/2021 Report. Volume I. New York: United Nations; 2022.

Bajs G. Doze zračenja na različitim digitalnim sustavima u radiologiji (Master’s thesis). Split: University of Split, Faculty of Health Sciences; 2026. (accessed 1 September 2026). Available at: https://repo.ozs.unist.hr/object/ozs:2089/FILE0

Bajs et al. Doze zračenja na različitim digitalnim sustavima u radiologiji